Interact CardioVasc Thorac Surg 2008;7:1131-1133. doi:10.1510/icvts.2008.186627
© 2008 European Association of Cardio-Thoracic Surgery

This Article
Right arrow Abstract Freely available
Right arrow Full Text (PDF)
Right arrow On-line Video
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Add to Personal Folders
Right arrow Download to citation manager
Right arrow Author home page(s):
Amihay Shinfeld
Ehud Raanani
Right arrow Permission Requests
Citing Articles
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Yuan, S.-M.
Right arrow Articles by Raanani, E.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Yuan, S.-M.
Right arrow Articles by Raanani, E.

Negative results - Valves

Displacement of the Amplatzer occluder device from the mitral paraprosthetic leak

Shi-Min Yuan*, Amihay Shinfeld and Ehud Raanani

Department of Cardiac and Thoracic Surgery, The Chaim Sheba Medical Center, Tel Hashomer 52621, Israel

Received 18 June 2008; received in revised form 18 July 2008; accepted 21 July 2008

Corresponding author. Tel.: +972 3 5302710; fax: +972 3 5302410.

E-mail address: shi_min_yuan{at}yahoo.com (S.-M. Yuan).


    Abstract
 Top
 Abstract
 1. Introduction
 2. Case report
 3. Discussion
 References
 
A critically ill patient had a major mitral paravalvular leak, failed device closure of the leak, and severe hemolytic anemia after a third mitral valve replacement. Echocardiography showed the Amplatzer occluder device completely displaced and swirled in the left atrial cavity. His condition was greatly improved after a fourth mitral valve replacement.

Key Words: Amplatzer occluder device; Complication; Hemolytic anemia; Mitral valve replacement; Paravalvular leak


    1. Introduction
 Top
 Abstract
 1. Introduction
 2. Case report
 3. Discussion
 References
 
As an alternative to surgical treatment, percutaneous transcatheter closure technique is typically employed in the management of congenital heart defects including atrial or ventricular septal defect, patent ductus arteriosus, and ruptured sinus of Valsalva [1, 2]. The practice of transcatheter closure of the paravalvular leak was started in 1987, by using the double-umbrella device developed by Rashkind and Cuaso [3]. Displacement of the Amplatzer occluder device is a rare complication, which occurred in 0.9–2.4% of the interventionally treated population with a congenital heart defect [4–6]. Only one case of displacement of the Amplatzer occluder device was reported subsequent to percutaneous closure of mitral paravalvular leak [7].


    2. Case report
 Top
 Abstract
 1. Introduction
 2. Case report
 3. Discussion
 References
 
A 65-year-old man, who had his third mitral valve replacement and subsequent paravalvular leak and failed device closure of the leak, recently developed severe hemolytic anemia, congestive heart failure, respiratory failure, renal failure, and klebsiella pneumonia bacteremia. He required prolonged ventilation through tracheostomy, multiple blood transfusion, and pacemaker for atrial fibrillation. He also had a history of diabetes, hypertension, ischemic heart disease, and glaucoma, etc. Laboratory tests (22 April 2008) showed that his hemoglobin was 8.7 g/dl (normal range, 14–18 g/dl), white blood cell count 22.94 K/µl (normal range, 4.8–10.8 K/µl), creatinine 2.81 mg/dl (normal range, 0.67–1.17 mg/dl), and lactate dehydrogenase 5133 U/l (normal range, 230–480 U/l). Echocardiography demonstrated mild wall motion abnormality, major mitral paravalvular leak, mild aortic valve stenosis, and mild tricuspid stenosis. Conservative treatments were not successful in improving his condition. He was referred to this department for surgical operation.

He was operated on under femorofemoral cardiopulmonary bypass on 4 June 2008. Preoperative echocardiography revealed a major posterior mitral paravalvular leak extending 0.8 cm in diameter, and the Amplatzer occluder device completely displaced and swirled in the left atrial cavity clockwisely (Fig. 1 and Videos 1 and 2). The left atrium was approached through the right atrium-atrial septum route. The tricuspid valve was regurgitant, the Amplatzer occluder device was located in the left atrium, and the mitral paravalvular leak was at the posterior portion. No signs of infection were noted in the mitral position. The Amplatzer occluder device was taken out (Fig. 2), and then the mitral valve was replaced with a 27 mm St. Jude Medical mechanical prosthesis (St. Jude Medical Inc., Minneapolis, MN, USA), and the tricuspid valve was repaired with a 32 mm Cosgrove–Edwards annuloplasty band (Edwards Lifesciences, Irvine, CA, USA). The cardiopulmonary bypass time was 228 min, and the cross-clamp time was 148 min.


Figure 1
View larger version (39K):
[in this window]
[in a new window]

 
Fig. 1. Preoperative transesophageal echocardiography revealed a major mitral paravalvular leak (lower arrow), and the Amplatzer occluder device completely displaced and swirled in the left atrium clockwisely (upper arrow).

 

Figure 3
View larger version (122K):
[in this window]
[in a new window]

 
Video 1. Intraoperative transesophageal echocardiography showing the Amplatzer occluder device completely displaced and swirled in the left atrium clockwisely.

 

Figure 4
View larger version (123K):
[in this window]
[in a new window]

 
Video 2. Color Doppler imaging of intraoperative transesophageal echocardiography showed the Amplatzer occluder device completely displaced and swirled in the left atrium clockwisely, and the major posterior mitral prosthetic leak.

 

Figure 2
View larger version (95K):
[in this window]
[in a new window]

 
Fig. 2. Operative photograph showing the Amplatzer occluder device was free in the left atrial cavity and was taken out via the right atrium-atrial septum approach.

 
The patient was hemodynamically stable and his condition was immediately improved after the operation. His hemoglobin was 9.5 g/dl, and his lactate dehydrogenase recovered to the normal limit. He was transferred to the Fifth Department of Internal Medicine for further convalescence on the 13th postoperative day.


    3. Discussion
 Top
 Abstract
 1. Introduction
 2. Case report
 3. Discussion
 References
 
Paravalvular leak is a relatively common complication of heart valve replacement. It may cause cardiac dysfunction and severe anemia, and lead the patient to a surgical intervention [8]. The advent of the transcatheter closure technique makes such patients with a high anticipated risk avoid redo operation [8]. Currently, Amplatzer occluder devices have been popularly accepted as a promising alternative to open heart surgery. However, significant residual regurgitation is sometimes noted, and repeated intervention is necessary [1]. The etiology of the residual paravalvular leak was primarily considered to be the irregular crescentric shape of the defect [1]. In the present patient, the suture dehiscence due to his poor conditions, including diabetes, may have been vulnerable to the paravalvular leak. Other complications of the occluder devices include arrhythmias, thromboembolism, device embolization, perforation, valve obstruction, and unstable position of the device [8].

Displacement of the Amplatzer occluder device is a rare complication of this interventional procedure. Only one case of occluder displacement after being deployed for mitral paravalvular leak was reported [7]. Size and location of the paravalvular leak, and presence of infection might be correlated with the stability of the occluder device. In addition, the device sitting on the atrial surface of the prosthesis was always moved by the eccentric regurgitant jet [7]. Such a complication requires surgical intervention to remove it from the intracardiac cavity or from the aorta [7, 9, 10]. Even though the present patient was critically ill and had developed a series of complications, surgical management seemed to be the only way for improving his condition.


    References
 Top
 Abstract
 1. Introduction
 2. Case report
 3. Discussion
 References
 

  1. Lasorda DM, Mohsin JC. Percutaneous closure of perivalvular mitral regurgitation with an Amplatzer occluder device in a patient with both prosthetic mitral and aortic valves. J Interv Cardiol 2008;21:190–195.[CrossRef][Medline]
  2. Schaeffler R, Sarikouch S, Peuster M. Transcatheter closure of a ruptured sinus of Valsalva aneurysm (RSVA) after aortic valve replacement using the Amplatzer muscular VSD occluder. Clin Res Cardiol 2007;96:904–906.[CrossRef][Medline]
  3. Hourihan M, Perry SB, Mandell VS, Keane JF, Rome JJ, Bittl JA, Lock JE. Transcatheter umbrella closure of valvular and paravalvular leaks. J Am Coll Cardiol 1992;20:1371–1377.[Abstract]
  4. Post MC, Van Deyk K, Budts W. Percutaneous closure of a patent foramen ovale: single-centre experience using different types of devices and mid-term outcome. Acta Cardiol 2005;60:515–519.[CrossRef][Medline]
  5. Lee CH, Kwok OH, Fan K, Chau E, Yip A, Chow WH. Transcatheter closure of atrial septal defect using Amplatzer septal occluder in Chinese adults. Catheter Cardiovasc Interv 2001;53:373–377.[CrossRef][Medline]
  6. Spies C, Timmermanns I, Schräder R. Transcatheter closure of secundum atrial septal defects in adults with the Amplatzer septal occluder: intermediate and long-term results. Clin Res Cardiol 2007;96:340–346.[CrossRef][Medline]
  7. Ussia GP, Scandura S, Calafiore AM, Mangiafico S, Meduri R, Galassi AR, Tamburino C. Images in cardiovascular medicine. Late device dislodgement after percutaneous closure of mitral prosthesis paravalvular leak with Amplatzer muscular ventricular septal defect occluder. Circulation 2007;115:e208–e210.[Free Full Text]
  8. Hein R, Wunderlich N, Robertson G, Wilson N, Sievert H. Catheter closure of paravalvular leak. EuroInv 2006;2:318–325.
  9. Tsilimingas NB, Reiter B, Kodolitsch YV, Münzel T, Meinertz T, Hofmann T. Surgical revision of an uncommonly dislocated self-expanding Amplatzer septal occluder device. Ann Thorac Surg 2004;78:686–687.[Abstract/Free Full Text]
  10. Bonatti J, Bonaros N, Müller S, Bartel T. Completely endoscopic removal of a dislocated Amplatzer atrial septal defect closure device. Interact Cardiovasc Thorac Surg 2008;7:130–132.[Abstract/Free Full Text]




This Article
Right arrow Abstract Freely available
Right arrow Full Text (PDF)
Right arrow On-line Video
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Add to Personal Folders
Right arrow Download to citation manager
Right arrow Author home page(s):
Amihay Shinfeld
Ehud Raanani
Right arrow Permission Requests
Citing Articles
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Yuan, S.-M.
Right arrow Articles by Raanani, E.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Yuan, S.-M.
Right arrow Articles by Raanani, E.


HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
ANN THORAC SURG ASIAN CARDIOVASC THORAC ANN EUR J CARDIOTHORAC SURG
J THORAC CARDIOVASC SURG ICVTS ALL CTSNet JOURNALS