ICVTS Click here for other ICVTS advertising opportunities
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


Interact CardioVasc Thorac Surg 2008;7:764-766. doi:10.1510/icvts.2008.177923
© 2008 European Association of Cardio-Thoracic Surgery

This Article
Right arrow Abstract Freely available
Right arrow Full Text (PDF)
Right arrow On-line Video
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Add to Personal Folders
Right arrow Download to citation manager
Right arrow Author home page(s):
Yukio Okazaki
Tsuyoshi Itoh
Right arrow Permission Requests
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Sakuragi, T.
Right arrow Articles by Itoh, T.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Sakuragi, T.
Right arrow Articles by Itoh, T.

Work in progress report - Thoracic general

Dramatic hemostasis of the transected pulmonary artery model using SOFT COAG electrosurgical output

Tohru Sakuragi*, Yukio Okazaki, Masahiro Mitsuoka and Tsuyoshi Itoh

Department of Thoracic and Cardiovascular Surgery, Faculty of Medicine, Saga University, Saga, Japan

Received 14 February 2008; received in revised form 2 June 2008; accepted 2 June 2008

*Corresponding author. 5-1-1 Nabeshima, Saga City 849-8501, Japan. Tel.: +81-952-34-2345; fax: +81-952-34-2061.

E-mail address: sakuragi{at}cc.saga-u.ac.jp (T. Sakuragi).


    Abstract
 Top
 Abstract
 1. Introduction
 2. Material and methods
 3. Results
 4. Discussion
 Acknowledgements
 References
 
We report the use of low-voltage, automatically regulated, electrosurgical coagulation to seal the bleeding from pulmonary arteries inadvertently during surgical intervention. Conventional electrosurgical coagulation uses high voltage, which generates intensive heat in the tissue. The heat results in carbonized eschar formation that can be easily broken by mechanical stress and lead to postoperative bleeding. SOFT COAG output automatically regulates the output voltage to a maximum of 200 Volts, preventing the generation of sparking. Thus, there is no formation of carbonized eschar. The instrument generates Joule heat alone in the tissue and the temperature rises to below boiling point, which effectively coagulates protein. Initial experiments, using a beagle model, clearly demonstrated the effectiveness and reliability of sealing both macroscopically and histopathologically. SOFT CAOG can be easily adopted both in open thoracotomy as well as in thoracoscopic procedures.

Key Words: Pulmonary artery; Bleeding; Hemostasis; SOFT COAG


    1. Introduction
 Top
 Abstract
 1. Introduction
 2. Material and methods
 3. Results
 4. Discussion
 Acknowledgements
 References
 
In anatomical pulmonary resection, such as lobectomy, we occasionally encounter accidental injury to the pulmonary artery. While the occurrence is rare, it can be life threatening [1]. If encountered during the thoracoscopic procedure, we must convert swiftly to a thoracotomy [2].

To be prepared for such a serious situation, we used the transected pulmonary artery model in beagle dogs to establish the validity of SOFT COAG output of a modern electrosurgical system (such as VIO300D, ERBE Elektromedizin GmbH, Germany) in closing.

Conventional electrosurgical units involve both Joule heating generated by the passing of electrical current through the electrical resistance of human tissue, which results in tissue temperatures well above the boiling point, and intensive heat generated by high voltage components, which forms carbonized eschar. Modern electrosurgical generators, such as the VIO300D (ERBE Elektromedizin GmbH, Germany) has a unique coagulation output mode called ‘SOFT COAG’, that automatically regulates its output voltage to stay below 200 Volts (V), causing the generation of Joule heat alone, and no tissue carbonization to result. Protein within the target tissue is effectively coagulated at temperatures between 70 and 80 °C due to the Joule heat generated in the tissue [3].


    2. Material and methods
 Top
 Abstract
 1. Introduction
 2. Material and methods
 3. Results
 4. Discussion
 Acknowledgements
 References
 
The SOFT COAG output of the VIO300D was evaluated to determine if it can seal transected pulmonary arteries in a beagle dog model. All of the animals involved in this study received humane care in compliance with the European Convention on Animal Care. The Animal Research Committee of the Saga University of Medicine approved all procedures. Beagle dogs (n=4) were anesthetized with an intravenous injection of pentobarbital sodium (15 mg/kg body weight) then endotracheally intubated. General anesthesia was maintained by inhalation of isoflurane under mechanical ventilation. The animal was placed in the left lateral position. The right chest was entered through the fifth intercostal space.

After the upper lobectomy, a pulmonary artery of the upper lobe was pinched up with forceps and transected at the root, which caused active bleeding (Videos 1, 2). After the temporary control of the bleeding with a finger, a ball electrode with a tip diameter of 3 mm (Fig. 1b) was placed against the bleeding point and moved in a circular motion. SOFT COAG was set at Effect 3 on the VIO300D. VIO's Effect represents the setting of power voltage. There are 8 levels of effects on the SOFT COAGV; Effect 1 is 55 V, and Effect 8, which is the max setting, is 190 V. It will not exceed over 200 V, even on max setting. Effect 3 is 88 V.


Figure 4
View larger version (127K):
[in this window]
[in a new window]

 
Video 1. After the upper lobectomy, a pulmonary artery was pinched up with forceps and it was transected at the root, inducing active bleeding. After temporary control of the bleeding with a finger, the tip of the ball electrode was applied to the bleeding point and moved in a circular motion. The bleeding point was closed immediately. The closed point did not bleed again even when scratched by a surgical instrument.

 

Figure 5
View larger version (109K):
[in this window]
[in a new window]

 
Video 2. The other three cases of the closure of the transected bleeding point; two of them well tolerated scratching after hemostasis.

 

Figure 1
View larger version (77K):
[in this window]
[in a new window]

 
Fig. 1. The electrosurgical system. (a) VIO300D generator (ERBE Electromedizin GmbH, Tuebingen, Germany). (b) Ball electrode with 3 mm tip diameter.

 

    3. Results
 Top
 Abstract
 1. Introduction
 2. Material and methods
 3. Results
 4. Discussion
 Acknowledgements
 References
 
Hemostasis was achieved almost immediately in one case (Video 1) while it took longer in the other three cases (Video 2). The sealed bleeding points did not bleed again even if rubbed vigorously by surgical instruments (Videos 1, 2). Histological examination showed that the transected lesion, identified in the defect of the elastic endothelial layer, was plugged with coagulated fibrin and surrounding connective tissue of the adventitia are coagulated showing fused collagen in the absence of tissue carbonization (Fig. 2). This phenomenon was estimated to be established by the momentary coagulation of the plasma fibrinogen into the fibrin and protein in the adventitia.


Figure 2
View larger version (129K):
[in this window]
[in a new window]

 
Fig. 2. Histological examination showed that the transected lesion, identified in the defect of the elastic endothelial layer (triangles), was plugged with coagulated fibrin and surrounding connective tissue of the adventitia are coagulated showing fused collagen in the absence of tissue carbonization (Fig. 2). This phenomenon was estimated to be established by the momentary coagulation of the plasma fibrinogen into the fibrin and protein in the adventitia.

 

    4. Discussion
 Top
 Abstract
 1. Introduction
 2. Material and methods
 3. Results
 4. Discussion
 Acknowledgements
 References
 
This is the first report of the hemostasis of active bleeding from the pulmonary artery using the SOFT COAG in an animal model. We were able to easily stop the bleeding from the root of the transected pulmonary artery by simply applying the tip of the ball-shaped electrode in a circular motion.

Transmitting low-frequency current through the human body causes severe injuries, such as electrochemical reactions and cardiac and respiratory malfunction [4]. Electric Surgical Unit, however, adopts high-frequency current (300 KHz ~ 1 MHz) to prevent these fatal accidents. The mechanism of conventional electrical coagulation is that electric current by set output flows into tissues, resulting in the rapid rise of the temperature within the tissues of the contact point. When cell fluid inside tissues reaches the boiling point and then turns dehydrated and dry, the flow of electric current is hindered, leading to voltage rise based on the principle (output=currentxvoltage). Above 200 Vp voltage produces sparks which incinerate the target tissues (voltage of conventional electrical coagulation is more than 1000 Vp) (Fig. 3a). The electrical coagulation with the high-temperature sparks is effective only for capillaries but not larger blood vessels due to the risk of their burst. Additionally, the intensive heat generated by the sparks form carbonized eschar. Tissue carbonization has a negative influence on the healing progress and leads to the risk of rebleeding by tissue peel-off. Modern electrosurgical generators, such as the VIO300D (ERBE Elektromedizin GmbH, Germany) has a unique coagulation output mode called ‘SOFT COAG’, that automatically regulates its output voltage to stay below 200 V (Fig. 3b), causing the generation of Joule heat alone, and no tissue carbonization to result. Protein within the target tissue is effectively coagulated at temperatures between 70 and 80 °C due to the Joule heat generated in the tissue [3].


Figure 3
View larger version (38K):
[in this window]
[in a new window]

 
Fig. 3. (a) Wave form of conventional coagulation: pulse-modulated sinusoidal alternating voltage Max. peak voltage: >1,000 Vp, (b) Wave form of SOFT COAG: unmodulated sinusoidal alternating voltage Max. peak voltage: <200 Vp.

 
Intense abrasion did not induce re-bleeding from the area sealed by this method. Subsequent histology of the treated area clearly demonstrated that coagulated fibrin and fused collagen led to the hemostasis. Carbonized eschar (not seen with this technique but, typical of the conventional electrosurgical coagulation), can be easily broken by such mechanical forces.

Bleeding from pulmonary artery during surgical procedures can be life threatening. While we have not yet used the SOFT COAG in a clinical setting, we believe that it will be useful in such cases. Even if bleeding could not be controlled completely, this method could buy precious time for a more permanent corrective action, such as suturing, to control the bleeding.


    Acknowledgements
 Top
 Abstract
 1. Introduction
 2. Material and methods
 3. Results
 4. Discussion
 Acknowledgements
 References
 
We thank Dr Edmund J. Miller, Director of Cardiopulmonary Research, The Feinstein Institute for Medical Research, for critical reviews.


    References
 Top
 Abstract
 1. Introduction
 2. Material and methods
 3. Results
 4. Discussion
 Acknowledgements
 References
 

  1. Daniels LJ, Balderson SS, Onaitis MW, D'Amico TA. Thoracoscopic lobectomy: a safe and effective strategy for patients with stage I lung cancer. Ann Thorac Surg 2002;74:860–864.[Abstract/Free Full Text]
  2. Watanabe A, Koyanagi T, Nakashima S, Higami T. How to clamp the main pulmonary artery during video-assisted thoracoscopic surgery lobectomy. Eur J Cardiothorac Surg 2007;31:129–131.[Abstract/Free Full Text]
  3. Reidenbach HD BG. Ancillary technology: electrocautery thermocoagulation and laser, Berlin, Springer.
  4. Lee R. Injury by electrical forces: pathophysiology, manifestations, and therpay. Curr Surg Prob 1997;34:677–764.[CrossRef]



This article has been cited by other articles:


Home page
ICVTSHome page
T. Sakuragi, H. Ohma, and H. Ohteki
Efficacy of SOFT COAG for intraoperative bleeding in thoracic surgery
Interactive CardioVascular and Thoracic Surgery, November 1, 2009; 9(5): 767 - 768.
[Abstract] [Full Text] [PDF]


This Article
Right arrow Abstract Freely available
Right arrow Full Text (PDF)
Right arrow On-line Video
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Add to Personal Folders
Right arrow Download to citation manager
Right arrow Author home page(s):
Yukio Okazaki
Tsuyoshi Itoh
Right arrow Permission Requests
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Sakuragi, T.
Right arrow Articles by Itoh, T.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Sakuragi, T.
Right arrow Articles by Itoh, T.


HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
ANN THORAC SURG ASIAN CARDIOVASC THORAC ANN EUR J CARDIOTHORAC SURG
J THORAC CARDIOVASC SURG ICVTS ALL CTSNet JOURNALS