Interact CardioVasc Thorac Surg 2007;6:425-429. doi:10.1510/icvts.2006.147256
© 2007 European Association of Cardio-Thoracic Surgery

This Article
Right arrow Abstract Freely available
Right arrow Full Text (PDF)
Right arrow On-line Video
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Add to Personal Folders
Right arrow Download to citation manager
Right arrow Author home page(s):
Dimitrios Dougenis
David Richens
Right arrow Permission Requests
Citing Articles
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Parissis, H.
Right arrow Articles by Richens, D.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Parissis, H.
Right arrow Articles by Richens, D.
Related Collections
Right arrow Cardiac - other

Work in progress report - Cardiac general

The way the intra-aortic balloon catheter moves within the aorta as a possible mechanism of balloon associated morbidity{star}

Haralambos Parissisa,*, Michael Leotsinidisb, Dimitrios Dougenisb and David Richensc

a Cardiothoracic Department, St James Hospital, D8, Dublin, Ireland
b University of Patras, Greece
c Nottingham City Hospital, UK

Received 21 December 2006; received in revised form 18 March 2007; accepted 20 March 2007

{star} Presented at the joint 20th Annual Meeting of the European Association for Cardio-thoracic Surgery and the 14th Annual Meeting of the European Society of Thoracic Surgeons, Stockholm, Sweden, September 10–13, 2006.

*Corresponding author. Tel.: +353-1-4103000; fax: +353-1-8265024.

E-mail address: hparissis{at}yahoo.co.uk (H. Parissis).


    Abstract
 Top
 Abstract
 1. Introduction
 2. Materials and methods
 3. Results
 4. Discussion
 Acknowledgements
 References
 
This study set off to investigate which mode of weaning of an intra-aortic balloon pump (IABP) produces more aortic trauma. With the use of a perfusion pump, an intact porcine aorta with an IABP in situ, was studied. Angioscopic images of the interior of the aorta were obtained. Whilst keeping steady blood pressure and flow, an ‘aortic impact score’ was calculated. Endoscopically there is a ‘whipping’ effect of the balloon shaft on the lateral aortic wall, which appears to be prominent in 1:3 mode. The aortic impact score at 0.5, 6 and 12 h during the experiments was: (1) When weaning by mode: a) 1:1 3.3±0.6, 4.0±1.0 and 4.3±0.6; b) 1:2 4.7±0.6, 6.7±0.6 and 7.0±0.0; c) 1:3 8.7±0.6, 11±1.0 and 11.7±0.6. (2) Weaning by augmentation: a) 75% 2.3±0.6, 2.7±0.6 and 3.0±0.0; b) 50% 1.3±0.6, 1.3±0.6 and 1.7±0.6. An increasing score was observed while weaning by mode (P<0.05). The 1:3 mode produces marked intimal disruption that worsens with time.

Key Words: IABP; Weaning of IABP; Complications of IABP


    1. Introduction
 Top
 Abstract
 1. Introduction
 2. Materials and methods
 3. Results
 4. Discussion
 Acknowledgements
 References
 
Morbidity related to intra-aortic balloon pump (IABP) treatment mainly occurs during insertion [1, 2]. Moreover, vascular complications during prolonged use of IABP have been quoted to occur in up to 8–10% of cases [3].

This study looks at the dynamic action of the IABP within the intact porcine aorta in an artificial circulation, in an attempt to delineate the movement of the balloon shaft within the aortic environment. This model also identifies modes of weaning of the IABP that could possibly lead to endothelial injury of the aorta.


    2. Materials and methods
 Top
 Abstract
 1. Introduction
 2. Materials and methods
 3. Results
 4. Discussion
 Acknowledgements
 References
 
Twenty-nine intact porcine aortae (animals 6–12 month old, weight 45–65 kg) were utilized, from animals at the slaughterhouse. The intrathoracic organs were excised ‘en block’.

Each aorta was rinsed with saline to remove blood clots from its lumen. All side branches were ligated (using minimal dissections) with the exception of the left subclavian and the common iliac vessels.

An artificial circulation was constructed. The circuit consisted of a standard 3/8 inch PVC perfusion tubing loop and a pulsatile dual chamber pump.

The pulsatile flow pump (Fig. 1a) consisted of a cylinder with a one-way valve on each side to direct flow. A pneumatic cylinder with a piston at its end drove the pump. The action of this fixture would work very much like the push–pull of a large syringe. The pneumatic cylinder was driven via a solenoid valve, which was controlled by a simple oscillator circuit. The system was triggered into action by pressurized air (60 psi), a heated reservoir and a 120/220 vacuum converter. The stroke volume of the pulsatile dual chamber pump was 50 ml (Data Scope Laboratories).


Figure 1
View larger version (65K):
[in this window]
[in a new window]

 
Fig. 1. (a) Schematic representation of the pulsatile pump (a cylinder with one way valve on each side to direct flow, driven by a pneumatic piston). (b) Schematic representation of the circuit (a porcine aorta was incorporated into the circuit with the inflow at the aortic valve and the outflow at the iliac artery). (c) Method of calculating the augmentation pressure. (d) Aorta with intimal fissuring, 12 h following counterpulsation Table 1. Aortic impact score.

 
The PVC loop was divided and an intact porcine aorta was incorporated into the circuit with the inflow at the aortic valve and the outflow at the right common iliac artery.

We utilized the intrathoracic and intra-abdominal organs ‘en block’. The supportive role of the tissues along the entire aorta and minimal tissue dissections, render an optimal anatomical model.

As a blood analogue Dextran 70 was used for perfusion. Pulsatile flow at rates of 50, 60 and 80 beats per min (bpm) were employed. Flow at an irregular rate (simulating atrial fibrillation) of 120 bpm was also tested.

Direct angioscopic images of the interior of the aorta were obtained using a rigid video endoscope (Olympus Keymed, camera OTV 54, light source CLV U20, Sony monitor) inserted via the left subclavian artery and snagged securely. An intra-aortic balloon (Datascope 9.5Fr) was then advanced under angioscopic control into the aorta via the left common iliac artery until its tip was just distal to the left subclavian artery (Fig. 1b). All the IABP used were appropriately sized and matched to the corresponding aorta.

Triggering of the balloon pump was achieved with a datascope ECG signal generator.

Blood pressure and flow stability was maintained throughout the experiments by continuous transfusion of the pump.

The traditional way of weaning has been to reduce the assist rate from 1:1 to 1:2 to 1:3 while maintaining diastolic augmentation at 100%.

Alternatively, weaning could be achieved by gradually decreasing the extent of diastolic augmentation while the rate is kept at 1:1. That can be achieved by calculating the augmentation pressure (Fig. 1c) from the presystolic point (C) to the peak of balloon augmentation (B).

This value is then decreased by 25% each time during two weaning phases. In order to decrease the diastolic augmentation by 25% the balloon volume should be decreased by 50%.

2.1. Protocol

With full augmentation and a pump rate of 80 bpm the first three sets of experiments were performed while the IABP was weaned by mode (in order to minimize bias, each experiment was repeated three times). During the next set of experiments the rate simulated atrial fibrillation of 120 bpm and the IABP was weaned by mode 1:1, 1:2.

During the last set of experiments the mode was kept 1:1 and the pump rate at 80 bpm. However, the augmentation was weaned to 75%, 50% and 25%, respectively.

Endoscopic monitoring and video recording was obtained throughout the experiments.

Biopsies at the proximal of the descending thoracic aorta were taken at 30 min, 6 h and 12 h following counterpulsation.

2.2. Endoscopic observation of the movement of the shaft of balloon

  1. Direct contact of the shaft of the deflated balloon on the aortic wall was scored subjectively as mild (score 1) moderate (score 2) or severe (score 3).
  2. Direct stroke of the shaft of the IABP on the aortic wall was scored as mild (score 1), moderate (score 2) or severe (score 3).
  3. Antegrade/retrograde movement of the shaft of the IABP was scored as mild (score 1), moderate (score 2) or severe (score 3).

2.3. Pathologic examination

Biopsies were taken throughout the experiment. Following mummification the specimen was examined under light microscopy (Zeiss Axiolab) with hematoxyline–iosine for tears at the inner layer.

The injury was scored:

Score 0 for no injury.

Mild (score 1) – when there were two or less endothelial disruptions per visual field (at magnification of 40). Moderate (score 2) – when there were 4–6 inner layer disruptions. Severe (score 3) – when there were 8–10 inner layer disruptions and dangerous (score 4) – when there were more than 10 inner layer disruptions per visual field.

The formula for ‘Aortic impact score’ (AIS) was derived as the Sum of scores:

  1. Direct contact of the shaft of IABP on the aortic wall
  2. Direct stroke of IABP on the aorta
  3. Back and forth shift of IABP
  4. Biopsy score at 30 min, 6 h and 12 h.


    3. Results
 Top
 Abstract
 1. Introduction
 2. Materials and methods
 3. Results
 4. Discussion
 Acknowledgements
 References
 
Perfusion could not be maintained in five out of the 29 aortae. The mean arterial pressure (MAP) recorded prior to intra-aortic balloon action was 45.4±8.32 mmHg. The MAP during intra-aortic balloon action on 1:1 was 88.6±9.18 mmHg and the peak augmentation pressure was 116.8±22.42 mmHg.

3.1. Angioscopic description

The entire length of the descending aorta was easily seen endoscopically. The balloon catheter did not occlude side branches of the aorta during augmentation.

Throughout the experiment a number of fine intimal flaps (Fig. 1d) were raised and visualized as 1–2 mm tears. The movement of the balloon within the aortic lumen appeared to be repetitive. The central axis of the balloon occupied a series of different positions during pumping (Video 1). During deflation the balloon was swept to one side of the aorta, by the circulation. This movement appeared to be enhanced during the prolonged phase of deflation associated with a 1:2 and especially with 1:3 mode (Fig. 2). The axis of the balloon re-obtained its central position at the end of inflation.


Figure 4
View larger version (75K):
[in this window]
[in a new window]

 
Video 1. Endoscopic view of the IABP, during 1:1 and 1:2 modes. The position of the central axis of the balloon relative to the aortic wall during deflation is constantly ‘around two o'clock’. This motion is prominent during 1:2 mode.

 

Figure 2
View larger version (86K):
[in this window]
[in a new window]

 
Fig. 2. The central axis of the balloon occupied a series of different positions during pumping. During deflation the balloon was swept to the same side of the aorta (whipping effect) by the circulation. This observation was mostly prominent during 1:3 mode.

 
The catheter also moved relative to the long axis of the aorta (Video 2). During inflation it advanced proximally towards the aortic valve. This position was held until deflation when it dropped back towards the aortic bifurcation as it was moving laterally towards the wall of the aorta (Fig. 3).


Figure 5
View larger version (83K):
[in this window]
[in a new window]

 
Video 2. Antegrade and retrograde movement of the IABP during 1:3 mode. This movement is less prominent during 1:1 mode.

 

Figure 3
View larger version (43K):
[in this window]
[in a new window]

 
Fig. 3. The catheter also moved relative to the long axis of the aorta (2–3 mm). During inflation it advanced proximally towards the aortic valve (antegrade movement). This position was held until deflation when it dropped back towards the aortic bifurcation (retrograde movement).

 
3.2. Weaning by mode vs. augmentation

Increasing AIS was observed with time, while weaning by mode (AIS 1:3>>AIS 1:2>AIS 1:1). The mean score was higher (see Table 1) at the group where the balloon was weaned by 1:3 mode (P<0.05).


View this table:
[in this window]
[in a new window]

 
Table 1 Aortic impact score (AIS) calculated at different time intervals and modes of weaning. A higher score was observed with time, while weaning by 1:3 mode. (AIS 1:3>>AIS 1:2 and AIS 1:3>>AIS augmentation) (P<0.05)

 
For all pair-wise comparisons between the two types of weaning (AIS 1:3>>AIS augmentation), the differences were statistically significant (P<0.05).

(ANOVA followed by Bonferroni post hoc test).

When the datascope ECG signal generator simulates atrial fibrillation, the AIS during various modes of weaning were low (4±2).


    4. Discussion
 Top
 Abstract
 1. Introduction
 2. Materials and methods
 3. Results
 4. Discussion
 Acknowledgements
 References
 
The relationship between the IABP and the intra-aortic environment has not been extensively investigated.

Wolvek [4] draws the attention on the hostile environment of the aging human aorta due to atherosclerotic plaques, or aneurysms. Under those conditions weaning the IABP by mode could be deleterious. Our group has shown in the past [5] that 1:3 mode leads to production of excess embolic material in human atherosclerotic aortas.

This study reveals that regardless of the mode of weaning, the balloon describes a repetitive motion. At maximal inflation the balloon lay centrally within the vessel. During deflation, however, the balloon was swept to one side by the passing circulation rather like ‘a sail in the wind’. Moreover, the deflationary position of the balloon was always observed to be at the same area of the aorta. This movement appears to be forceful (whipping effect) during 1:3 mode.

The movement of the central axis of the balloon catheter could, therefore, be described as circular. During inflation the balloon finds a central position within the lumen and advances by 2–3 mm toward the aortic valve. During deflation the balloon falls back towards the aortic bifurcation by a similar distance and is swept sideways striking the posterolateral wall of the vessel. This position is held longer during 1:2 mode due to the prolong period of deflation. We can assume that this may be a mechanism of balloon perforation, if the catheter is constantly striking and then rubbing alongside an irregular atheromatous area.

Direct contact between the catheter and arterial wall may also be a mechanism of intimal insult and fissuring of plaque resulting on embolization. The intra-aortic balloon catheter typically cycles at 144,000 beats every 24 h [6]. The increments of repetitive pulse pressure applied on the aortic wall during each cycle may be significant enough to produce excess insult, especially in a calcified inelastic atheromatous aorta.

During this study we have constantly observed high AIS in 1:3 mode. This supports our hypothesis that 1:3 mode has a potential deleterious impact on the aortic wall.

In contrary, atrial fibrillation was not associated with high AIS. There were no differences between 1:1 and 1:2 mode of weaning. This reflects our endoscopic observation that during periods of atrial fibrillation, the balloon shaft movement appears to be weak with the ‘whipping effect’ being absent.

The prolong duration of IABP therapy was associated with higher AIS, in all modes of weaning.

The impact of prolong IABP treatment on the development of complications, is controversial. Duration of use was the only balloon-related variable that approached significance (P=0.051) in the published studies of Funk et al. [7]. Contrary, Alderman et al. [8] and Gottlieb et al. [9] identified incremental risk factors for the development of vascular complications as being: female sex, peripheral vascular disease and diabetes.

Freed and associates [10] analyzed 733 cases with specific interest upon the complications during prolonged circulatory support with IABP. Vascular problems were frequent in prolonged assisted patients.

Similar conclusions are reported by Iverson et al. [11]. They concluded that patients who required balloon support for more than 60 h had 1.5 times the complication rate compared with those requiring support for less (32% vs. 21%).

In our study the highest score was observed during prolong IABP weaning with 1:3 mode.

There are weaknesses in this study due to:

  1. The potential bias from the observational nature of the study.
  2. There were difficulties throughout the experiments in maintaining a constant blood pressure and flow.
  3. There was a subjective factor in some of the variables used in order to calculate the AIS.

Limb ischemia during treatment with IABP is multi-factorial and mainly reflects the atherosclerotic status of the descending aorta and the difficulty of dealing with a severely atherosclerotic femoral artery. Our study suggests that ‘weaning by mode’ could be a risk factor for adverse vascular outcome.

In conclusion, direct stroke effect and sliding motions of the balloon shaft against the aortic wall (mainly during 1:2 and 1:3 modes) may be involved in balloon-associated morbidity. The clinical implication of that would be avoidance of those modes during weaning. A clinical trial remains to confirm the latter observation.


    Acknowledgements
 Top
 Abstract
 1. Introduction
 2. Materials and methods
 3. Results
 4. Discussion
 Acknowledgements
 References
 
The authors gratefully acknowledge the contribution of Paul Quinlan for the photography.


    References
 Top
 Abstract
 1. Introduction
 2. Materials and methods
 3. Results
 4. Discussion
 Acknowledgements
 References
 

  1. McCabe JC, Abel RM, Subramanian VA, Gay W. Complications of intra-aortic balloon insertion and counterpulsation. Circulation 1978; 57:769.[Medline]
  2. Alvarez JM, Gates R, Rowe D, Brady PW. Complications from intraaortic balloon counterpulsation: a review of 303 cardiac surgical patients. Eur J Cardio-Thorac Surg 1992; 6:530–535.[Abstract]
  3. Goldman BS, Hill TS, Rosenthal GA. Complications associated with use of the intraaortic balloon pump. Can J Surg 1982; 25:153.[Medline]
  4. Wolvek S. The hostile environment of the aging human aorta and the smaller patient – their implications for the intra-aortic balloon. Perfusion 1994; 9:87–94.[Abstract/Free Full Text]
  5. Tierney G, Parissis H, Baker M, Austin D, Clelland C, Richens D. An experimental study of intra aortic balloon pumping within the intact human aorta. Eur J CardioThorac Surg 1997; 12:486–493.[Abstract]
  6. Pace PD, Tilney NL, Lesch M, Couch NP. Peripheral arterial complications of intra-aortic balloon counterpulsation. Surgery 1977; 82:685–688.[Medline]
  7. Funk M, Gleason J, Foell D. Lower limb ischemia related to use of the intraaortic balloon pump. Heart Lung 1989; 18:542.[Medline]
  8. Alderman JD, Gabliani GI, McCabe CH. Incidence and management of limb ischemia with percutaneous wire quided intraaortic balloon catheters. J Am Coll Cardiol 1987; 9:524–530.[Abstract]
  9. Gottlieb SO, Brinker JA, Borkon AM. Identification of patients at high risk for complications of intraaortic balloon counterpulsation: a multivariate risk factor analysis. Am J Cardiol 1984; 53:1135.[CrossRef][Medline]
  10. Freed P, Wasfie T, Zado B, Kantrowitz A. Intraaortic balloon pumping for prolonged circulatory support. Am J Cardiol 1988; 61:554.[CrossRef][Medline]
  11. Iverson L, Herfindahl G, Ecker R, Young N, Ennix C, Lee J, Dunning C, Whisenant A, May I. Am J Surg 1987; 154:99.[Medline]




This Article
Right arrow Abstract Freely available
Right arrow Full Text (PDF)
Right arrow On-line Video
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Add to Personal Folders
Right arrow Download to citation manager
Right arrow Author home page(s):
Dimitrios Dougenis
David Richens
Right arrow Permission Requests
Citing Articles
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Parissis, H.
Right arrow Articles by Richens, D.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Parissis, H.
Right arrow Articles by Richens, D.
Related Collections
Right arrow Cardiac - other


HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
ANN THORAC SURG ASIAN CARDIOVASC THORAC ANN EUR J CARDIOTHORAC SURG
J THORAC CARDIOVASC SURG ICVTS ALL CTSNet JOURNALS