ICVTS Click here for other ICVTS advertising opportunities
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


Interact CardioVasc Thorac Surg 2006;5:301-302. doi:10.1510/icvts.2005.127423
© 2006 European Association of Cardio-Thoracic Surgery

This Article
Right arrow Abstract Freely available
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Add to Personal Folders
Right arrow Download to citation manager
Right arrow Author home page(s):
Hsi-Yu Yu
Yih-Sharng Chen
Right arrow Permission Requests
Citing Articles
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Lee, K.-K.
Right arrow Articles by Chen, Y.-S.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Lee, K.-K.
Right arrow Articles by Chen, Y.-S.
Related Collections
Right arrow Cardiac - other
Right arrow Congestive Heart Failure

New ideas - Cardiac general

Hand-made sizing balloon in surgical ventricular restoration

Kun-Kuang Leea, Chih-Hsien Wanga, Hsi-Yu Yua and Yih-Sharng Chena,b,*

a Department of Surgery, National Taiwan University Hospital, Taipei, Taiwan
b Department of Surgery, National Taiwan University Hospital Yun-Lin Branch, 7 Chung-Shan S. Rd., Taipei, Taiwan

Received 23 December 2005; received in revised form 16 February 2006; accepted 17 February 2006

*Corresponding author. Fax: +886-2-23958747.

E-mail address: yschen11{at}yahoo.com.tw (Y.-S. Chen).


    Abstract
 Top
 Abstract
 1. Introduction
 2. Technique
 3. Comment
 References
 
Left ventricular size and shape are considered as a key issue in surgical ventricular restoration procedure. A commercially preshaped balloon could provide a guide to perform the procedure. We developed a new, reliable and simple method with simple material and without additional sterilization to achieve the similar goal.

Key Words: Heart failure; Surgical restoration; Sizing balloon


    1. Introduction
 Top
 Abstract
 1. Introduction
 2. Technique
 3. Comment
 References
 
Postinfarction left ventricular remodeling is characterized by chamber dilatation and abnormal shape leading to systolic and diastolic dysfunction, and then heart failure syndrome. Surgical ventricular restoration (SVR) is designed to diminish the dilated ventricular volume and restore the shape. If the residual ventricular volume is too small, the pathophysiology of restrictive cardiomyopathy will result in a catastrophic event; if the residual ventricular volume is too large, the result will be the same condition as the preoperative status of heart failure or even worse.

To diminish the risk, Dor introduced the use of an intraventricular balloon to guide the volume reduction [1,2]. The volume 60 ml/m2 is chosen for the balloon size after study. More recently, a preshaped elliptical balloon (Chase Medical, Dallas, TX) is used to size and configure the ventricle, achieving a more normal short axis/long axis ratio and it gives us the correct position of the new apex.

We designed a new method with a hand-made sizing balloon to achieve this goal.


    2. Technique
 Top
 Abstract
 1. Introduction
 2. Technique
 3. Comment
 References
 
We cut off the basal part of the middle finger and its tip of a 71/2 size plastic surgical glove (Fig. 1). Then we make a silk tie in the base and its tip is tied to a connector with another silk. The connector is connected to a three-way lock (Fig. 2). During the operation, we insert the balloon into the ventricle and inflate it with the desired volume of normal saline using a 20-ml syringe to size the ventricle. Also, the glove's finger tip may be used to point the direction for the new apex; while the glove's base may be used to guide the basal part of the left ventricle. When the patch closure is nearly completed, we deflate the balloon and remove it. The following procedure is continued.


Figure 1
View larger version (104K):
[in this window]
[in a new window]
 
Fig. 1. The basal part of the middle finger and its tip of a 71/2 size surgical plastic glove are cut off.

 

Figure 2
View larger version (114K):
[in this window]
[in a new window]
 
Fig. 2. The basal part of the middle finger of a 71/2 size surgical plastic glove is tied by a silk and its tip is tied by another silk to a connector, which is connected to a three-way lock. Normal saline is injected into the hand-made balloon by a 20-ml syringe.

 

    3. Comment
 Top
 Abstract
 1. Introduction
 2. Technique
 3. Comment
 References
 
Whether a sizing balloon is mandatory for the SVR still requires further investigation, it is not cost effective to prepare every size of the designed balloon commercially available. The hand-made sizing balloon using a surgical plastic glove is cheap and easily available in the operating room. The procedure is also easy to perform and reproducible. After inflation by normal saline, it has a nearly elliptical shape and its finger tip can be used as a guide for the new apex. Its obtuse part can be used as a guide of the left ventricular base. The features are almost the same as the commercialized preshaped sizing balloon. One advantage of this hand-made balloon is that we don't have to remove it until the last stitches are placed because it can be completely deflated. And this is not possible for the rigid commercially available preshaped balloon. But, on the contrary, the hand-made balloon may be compressed and the dimensions changed despite the same quantity of saline present because it is not as rigid as the commercialized preshaped sizing balloon.

We have performed SVR in 10 patients in the past 2 years and two of them were applied with this modified method. The shape of the ventricle could be maintained with this method and the hand-made balloon is much easier to extract after deflation.

In conclusion, the hand-made sizing balloon may be a cheap and practical alternative choice in surgical ventricular restoration.


    References
 Top
 Abstract
 1. Introduction
 2. Technique
 3. Comment
 References
 

  1. Dor V, Saab M, Coste P, Komaszewska M, Montiglio F. Left ventricular aneurysm: new surgical approach. Thorac Cardiovasc Surg 1989; 37:11–19.[Medline]
  2. Dor V. Surgery for left ventricular aneurysm. Curr Opin Cardiol 1990; 5:773–780.[Medline]




This Article
Right arrow Abstract Freely available
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Add to Personal Folders
Right arrow Download to citation manager
Right arrow Author home page(s):
Hsi-Yu Yu
Yih-Sharng Chen
Right arrow Permission Requests
Citing Articles
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Lee, K.-K.
Right arrow Articles by Chen, Y.-S.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Lee, K.-K.
Right arrow Articles by Chen, Y.-S.
Related Collections
Right arrow Cardiac - other
Right arrow Congestive Heart Failure


HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
ANN THORAC SURG ASIAN CARDIOVASC THORAC ANN EUR J CARDIOTHORAC SURG
J THORAC CARDIOVASC SURG ICVTS ALL CTSNet JOURNALS