ICVTS Click here for other ICVTS advertising opportunities
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


Interact CardioVasc Thorac Surg 2009;8:579-580. doi:10.1510/icvts.2009.202077
© 2009 European Association of Cardio-Thoracic Surgery

This Article
Right arrow Full Text
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Add to Personal Folders
Right arrow Download to citation manager
Right arrow Author home page(s):
Martin Strüber
Right arrow Permission Requests
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Oswald, H.
Right arrow Articles by Gardiwal, A.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Oswald, H.
Right arrow Articles by Gardiwal, A.

Brief communication - Arrhythmia

Implantable defibrillator with left ventricular assist device compatibility

Hanno Oswalda,*, Gunnar Kleina, Martin Strüberb and Ajmal Gardiwala

a Department of Cardiovascular Medicine, Hannover Medical School, Carl-Neuberg-Str. 1, 30625 Hannover, Germany
b Department of Cardiothoracic, Transplantation and Vascular Surgery, Hannover Medical School, Hannover, Germany

*Corresponding author. Tel.: +49 511 532 3817; fax: +49 511 532 8475.

E-mail address: hanno_oswald{at}gmx.de (H. Oswald).

Implanted defibrillator (ICD) and HeartMate-II (Thoratec, Pleasanton, CA, USA) left ventricular assist device (LVAD) incompatibilities have been reported. In order to show which ICD would function regularly in Heart Mate II recipients, we conducted a systematic review of 39 consecutive patients with implanted ICD and HeartMate-II from our institution's LVAD registry. Forty-six ICDs were identified without device interactions in most ICD devices. However, loss of ICD to programmer telemetry was observed in four patients with V-193, V-196 and V-350 (St Jude Medical, Sunnyvale, CA, USA) and one patient with an Ovatio DR (Sorin, Milan, Italy) requiring ICD replacement. The majority of all investigated ICDs operated without LVAD to ICD interaction. However, surgeons implanting the Heart Mate II in ICD patients should be aware of possible interactions and incompatible devices should not be recommended in HeartMate-II recipients.

Key Words: ICD; Implantable defibrillator; LVAD; Left ventricular assist device




This article has been cited by other articles:


Home page
J Am Coll CardiolHome page
S. R. Wilson, M. M. Givertz, G. C. Stewart, and G. H. Mudge Jr
Ventricular assist devices the challenges of outpatient management.
J. Am. Coll. Cardiol., October 27, 2009; 54(18): 1647 - 1659.