Interact CardioVasc Thorac Surg 2008;7:1131-1133. doi:10.1510/icvts.2008.186627 © 2008 European Association of Cardio-Thoracic Surgery
Displacement of the Amplatzer occluder device from the mitral paraprosthetic leakDepartment of Cardiac and Thoracic Surgery, The Chaim Sheba Medical Center, Tel Hashomer 52621, Israel
Corresponding author. Tel.: +972 3 5302710; fax: +972 3 5302410. A critically ill patient had a major mitral paravalvular leak, failed device closure of the leak, and severe hemolytic anemia after a third mitral valve replacement. Echocardiography showed the Amplatzer occluder device completely displaced and swirled in the left atrial cavity. His condition was greatly improved after a fourth mitral valve replacement.
Key Words: Amplatzer occluder device; Complication; Hemolytic anemia; Mitral valve replacement; Paravalvular leak
|