ICVTS Click here to locate an Ethicon representative
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


Interact CardioVasc Thorac Surg 2008;7:750-754. doi:10.1510/icvts.2008.182469
© 2008 European Association of Cardio-Thoracic Surgery

This Article
Right arrow Full Text
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Add to Personal Folders
Right arrow Download to citation manager
Right arrow Author home page(s):
Tomaso Bottio
Giulio Rizzoli
Gino Gerosa
Right arrow Permission Requests
Google Scholar
Right arrow Articles by Bottio, T.
Right arrow Articles by Gerosa, G.
PubMed
Right arrow PubMed Citation
Right arrow Articles by Bottio, T.
Right arrow Articles by Gerosa, G.

Work in progress report - Valves

The changing spectrum of bioprostheses hydrodynamic performance: considerations on in-vitro tests

Tomaso Bottio*, Vincenzo Tarzia, Giulio Rizzoli and Gino Gerosa

Cardiovascular Institute, University of Padova, Via Giustiniani 2, 36100 Padova, Italy

*Corresponding author. Tel.: +39-049-8212408; fax: +39-049-8212409.

E-mail address: tbottio{at}gmail.com (T. Bottio).

The aims of the present study were to compare hydrodynamics of three pericardial and two porcine valves while performing at different stroke volume (SV) and increasing pulse rate (PR). Carpentier-Edwards Magna-21 (CEM), Sorin Soprano-20 (SS), Mitroflow-23 (MF), SJM-Biocor-Epic-Supra-21 (SJME), and Medtronic Mosaic Ultra-23 (MMU) were tested in the aortic chamber (23-mm in diameter) of the Sheffield-Pulse-Duplicator. The tests were carried out at increasing pulse-rate and at each pulse-rate the valve was tested at different SV. CEM and MF showed significantly lower gradients than porcine valves and SS. Transvalvular gradients were unrelated to PR showing a constant value with increasing PR. While SJME valve showed the lowest regurgitant volume, on the contrary CEM showed the highest. At increasing SV, effective-orifice-area (EOA) observed with CEM was significantly larger than with the other tested valves, even though at SV 60 ml MF was comparable and at SV 65 ml significantly better. SS, SJME and MMU showed comparable EOAs with bigger area at increasing PR. The latter relation was reversed for CEM and MF. Our results show that CEM and MF have shown significantly better in-vitro hydrodynamics in comparison with their porcine counterpart and SS. Nevertheless, at increasing pulse rate, porcine tissue valves and SS may guarantee higher EOA values.

Key Words: Sheffield Pulse Duplicator; Supra-annular tissue valves; In-vitro tests; Patient prosthesis mismatch